Online TN Analyzer, HATN-2000, is a measuring equipment that quantifies total nitrogen by using the UV absorption spectrophotometry method. By oxidizing various nitrogenous compounds into nitrate nitrogen through oxidative degradation, it is quantified by using the UV absorption spectrophotometry. HATN-2000 has a simple configuration that has adopted the 7-port control valve to enable inflow of fixed quantity of sample and reagent for the purpose of accurate measurement. It has less clogging due to the use of large tubes, and it can automatically check the supply of sample and reagent.
Characteristics
HATN-2000
-
Simple structure:7-port control valve
-
Improved measurement accuracy
-
Compliance with the Standard Water Pollutant Testing Method
-
Economical maintenance expenses
-
Status condition display
-
Functions that separately discharges hazardous
waste liquid and general liquid (Option)
Principle
- In the pretreatment step, by applying alkaline potassium persulfate and high heat to the sample, organic and inorganic nitrogen compound is turned into nitrate ions by using oxidative degradation. After cooling, the acidity is lowered and absorbance is measured to quantify the total nitrogen. The absorbance of nitrate ion is measured at 220nm, and the interference factor is compensated at 275nm.
Specification
-
-
- Measurement item
-
Total Nitrogen (TN)
-
- Measurement method
-
UV Absorption photometry
-
- Measurement range
-
0~100mg/L (Adjustable)
-
- Wavelength
-
220nm (275nm for compensation)
-
- Measurement interval
-
60 minutes
-
- Accuracy
-
±2% F.S.
-
- Reproducibility
-
±2% F.S.
-
- Linearity
-
±3% F.S.
-
- Zero/Span drift
-
±2% F.S.
-
- Oxidation method
-
120℃ persulfate thermal oxidation
-
- Calibration
-
Auto/manual calibration by standard solution
-
- Sample/Reagent measurement method
-
Photo sensor pulse control
-
-
-
- Sample/Reagent distribution method
-
7-port distribution control method (Fixed quantity injection with multi-channel valve system)
-
- Operation system
-
Windows 7 (CPU process) or more
-
- Output
-
TCP/IP, DC 4~20mA, RS-232C
-
- Status display
-
Normal operation, on calibration, error
-
- Display
-
Color LCD 10.4″ (Touch screen)
-
- Power
-
AC 220V, 50/60Hz
-
- Power consumption
-
Average 150W(Maximum 450W)
-
- Voltage regulation
-
3% or less
-
- Data storage capacity
-
HDD 100G or more
-
- Data output
-
USB terminal
-
- Enclosure / Size
-
Steelã€IP55 / 600(W) â…¹ 662(D) â…¹ 1700(H) mm
-
- Operating humidity
-
RH 95% or less
-
Application
Application Field
-
Surface water monitoring
- River, lake, river water quality
- Waterworks facilities (intake station and Water plant, etc.)
-
Water quality TMS
- Sewage treatment plant, Wastewater treatment plant
- Wastewater discharging facility, Rural sewage treatment facility, etc.
-
Process control, etc.
- STP process management,
- IWWTP process management, etc.
Installation Pictures
-
Yongin Respia
-
Naepo