Water quality monitoring of Online TOX/BOD System – HABT-3000

TOX/BOD System, HABT-3000, is a system designed to measure BOD and monitor toxic substances by using MFC as a sensor. The MFC used in HABT-3000 enables measurement of BOD on diverse organic concentration and properties by using complex microbes that live in the field, and at the same time, it is a model for real-time measurement that can be applied to detection of various toxic substances within a sample.

Characteristics

HABT-3000
  • Simultaneous Measurement of Toxic Substance (TOX) / BOD
    • InnovationSimultaneous measurement of organic concentration and toxic substances by using MFC
    • GreenNo emission of harmful chemicals
      Safe for the environment by using microbes living in the ecosystem
  • Rapidity & Sensitivity
    • RapidityReaction within 10 minutes
    • SensitivitySuperior sensitivity compare to other technologies through the use of microbes
      Reaction to diverse harmful substances
  • Convenience
    • ConvenienceDoesn’t require cultivation or supply of test organisms
      Easy and convenient operation
    • CostDoesn’t require test organism expenditure or replacement
      Doesn’t require cultivation and management personnel

Principle

  • When the sample flows into the sensor, the organic matter within the sample oxidizes by the microbes and electric signal is generated. Because the amount of electricity generated is proportional to the concentration of organic matter, the BOD of the sample can be measured. On the other hand, because the organic matter dies or the metabolism is blocked due to reduction of current production value when toxic substances flow in, the inflow of toxic substance can be detected..

Specification

    • Measurement item
      Toxicity (TOX) | Biochemical Oxygen Demand (BOD)
    • Measurement method
      Microbial fuel cell(MFC) using electrochemically active bacteria(EAB)
    • Working electrode
      Microbial Fuel Cell (MFC)
    • Measurement range
      -100~+100% | 0~30mg/L
    • Measurement interval
      10 minutes (Adjustable) | 10 minutes (Adjustable)
    • Accuracy
      – | ±5% F.S.
    • Reproducibility
      – | ±5% F.S.
    • Linearity
      – | ±5% F.S.
    • Zero/Span drift
      – | ±5% F.S.
    • Operation system
      Windows 7 (CPU process) or more
    • Output
      TCP/IP, DC 4~20mA, RS-232C
    • Status display
      Normal operation, on calibration, error
    • Display
      Color LED 10.4″(Touch screen)
    • Power
      AC 220V、50/60Hz
    • Voltage regulation
      3% or less
    • Data storage capacity
      HDD 100G or more
    • Data output
      USB terminal
    • Enclosure / Size
      Steel , 1.5T (Except paint) / 645(W) â…¹ 439(D) â…¹ 1600(H) mm
    • Automatic thermostat
      Heater by PI control
    • Operating humidity
      RH 95% or less

Application

Application Field
  • Surface water monitoring
    • River, lake, river water quality
    • Waterworks facilities (intake station and Water plant, etc.)
  • Water quality TMS
    • Sewage treatment plant, Wastewater treatment plant
    • Wastewater discharging facility, Rural sewage treatment facility, etc.
  • Process control, etc.
    • STP process management,
    • IWWTP process management, etc.

Leave a Reply