Water quality monitoring system of Online TN Analyzer – HATN-2000

Online TN Analyzer, HATN-2000, is a measuring equipment that quantifies total nitrogen by using the UV absorption spectrophotometry method. By oxidizing various nitrogenous compounds into nitrate nitrogen through oxidative degradation, it is quantified by using the UV absorption spectrophotometry. HATN-2000 has a simple configuration that has adopted the 7-port control valve to enable inflow of fixed quantity of sample and reagent for the purpose of accurate measurement. It has less clogging due to the use of large tubes, and it can automatically check the supply of sample and reagent.

Characteristics

HATN-2000
  • Simple structure:7-port control valve
  • Improved measurement accuracy
  • Compliance with the Standard Water Pollutant Testing Method
  • Economical maintenance expenses
  • Status condition display
  • Functions that separately discharges hazardous
    waste liquid and general liquid (Option)

Principle

  • In the pretreatment step, by applying alkaline potassium persulfate and high heat to the sample, organic and inorganic nitrogen compound is turned into nitrate ions by using oxidative degradation. After cooling, the acidity is lowered and absorbance is measured to quantify the total nitrogen. The absorbance of nitrate ion is measured at 220nm, and the interference factor is compensated at 275nm.

Specification

    • Measurement item
      Total Nitrogen (TN)
    • Measurement method
      UV Absorption photometry
    • Measurement range
      0~100mg/L (Adjustable)
    • Wavelength
      220nm (275nm for compensation)
    • Measurement interval
      60 minutes
    • Accuracy
      ±2% F.S.
    • Reproducibility
      ±2% F.S.
    • Linearity
      ±3% F.S.
    • Zero/Span drift
      ±2% F.S.
    • Oxidation method
      120℃ persulfate thermal oxidation
    • Calibration
      Auto/manual calibration by standard solution
    • Sample/Reagent measurement method
      Photo sensor pulse control
    • Sample/Reagent distribution method
      7-port distribution control method (Fixed quantity injection with multi-channel valve system)
    • Operation system
      Windows 7 (CPU process) or more
    • Output
      TCP/IP, DC 4~20mA, RS-232C
    • Status display
      Normal operation, on calibration, error
    • Display
      Color LCD 10.4″ (Touch screen)
    • Power
      AC 220V, 50/60Hz
    • Power consumption
      Average 150W(Maximum 450W)
    • Voltage regulation
      3% or less
    • Data storage capacity
      HDD 100G or more
    • Data output
      USB terminal
    • Enclosure / Size
      Steel、IP55 / 600(W) ⅹ 662(D) ⅹ 1700(H) mm
    • Operating humidity
      RH 95% or less

Application

Application Field
  • Surface water monitoring
    • River, lake, river water quality
    • Waterworks facilities (intake station and Water plant, etc.)
  • Water quality TMS
    • Sewage treatment plant, Wastewater treatment plant
    • Wastewater discharging facility, Rural sewage treatment facility, etc.
  • Process control, etc.
    • STP process management,
    • IWWTP process management, etc.
Installation Pictures
  • Yongin Respia
  • Naepo

Leave a Reply